GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5 VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures 8 - 203
8-8-7-1 Preparations (cont’d)
4.) Remove five OP Panel Knobs, see Figure 8-167.
5.) Remove the Operator Panel, Upper.
6.) Remove the Operator Panel, Lower.
7.) Remove the Left Side Cover.
8.) Remove the Top Cover.
9.) Remove the Main Cable Cover.
8-8-7-2 Main Cable removal procedure
Three cable connectors must be disconnected in the Operator Panel area:
1.) The cable with the D-SUB connector was disconnected from the Controller Board when you
removed the Operator Panel, Upper.
2.) Disconnect the HDMI cable from the Bulkhead.
3.) Disconnect the cable from the Z switch (Up/Down control).
4.) If not already done, unscrew the fixing screws on the Ground (GND) Clamp on the Bulkhead
Bracket, so you can release the cables from the clamp.
Two cables are connected to the XY (Frog) brakes. They are routed to connectors inside the rear XY
(Frog) legs. To disconnect these cables, follow this procedure:
1.) Remove the covers on the rear XY (Frog) legs.
2.) Disconnect the cables.
Two (2x) cables (from the Main Cable) are connected to the XYZ Control box.
Disconnect the two cables.
The remaining two cables originate on the I/O at the BEP.
1.) Disconnect the DVI cable from J22.
2.) Disconnect the D-SUB from J21.
Figure 8-209 Remove five knobs

Table of Contents