Chapter 8 - Replacement procedures 8 - 257
8-11-3-1 Preparations (cont’d)
3.) If possible, wipe the HDD partitions as described here: Reload the system software from DVD.
Select A: to overwrite all content on HDD.
4.) Power down the VIVID E9.
5.) Disconnect the Mains Power Cable from the wall outlet.
6.) Disconnect all external cables (probes, ECG, physio, ethernet, audio, video, USB).
7.) Remove the Left Side Cover.
8.) Remove the BEP Cover.
a.) Loosen the BEP Cover’s fixing screw(s).
- On BEP6 there are two finger screws on the top of the BEP Cover.
- On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
- On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the BEP Cover away from the BEP.
c.) Set the BEP Cover aside.