8 - 296 Section 8-13 - I/O Modules replacement
8-13-1-1 Preparations (cont’d)
4.) Remove Left Side cover.
5.) Remove the BEP Cover.
a.) Loosen the BEP Cover’s fixing screw(s).
- On BEP6 there are two finger screws on the top of the BEP Cover.
- On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
- On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the BEP Cover away from the BEP.
c.) Set the BEP Cover aside.
6.) BEP6 Only: Remove the BEP6 Power Board Assembly.
8-13-1-2 Patient I/O module removal procedure
Follow these steps to remove the Patient I/O module:
1.) Inside the BEP there is an opening for the Power and USB cables to the Patient I/O module.
Disconnect the Power cable connector and the USB cable connector from the Patient I/O module.
2.) Remove one screw inside the BEP.
3.) Remove the two fastening screws on the Patient I/O module’s mounting bracket.
4.) Pull the Patient I/O module sideways, out of the frame.
8-13-1-3 Patient I/O installation procedure
Follow these steps to install the Patient I/O:
1.) Align the Patient I/O with the opening and push it into its final position.
2.) Connect the Power and USB cables (inside the BEP).
3.) BEP5 only: Install one screw inside the BEP.
4.) Install two fastening screws on the mounting bracket.
5.) BEP6 Only: Install the BEP6 Power Board Assembly.
Figure 8-248 Power and USB connectors (rear side of Patient I/O)
POWER
CONNECTOR
USB2 SIGNAL
CONNECTOR