10 - 16 Section 10-5 - System maintenance
10-5-6-5 Cleaning TEE probes
To clean the handle:
• Use a cloth or towel only lightly moistened with cleaner/ disinfectant (as recommended in TEE
Probe Care Card).
Avoid too much moisture on the cloth/ towel.
• To remove residual chemicals wipe the handle with a cloth or towel lightly moistened with clean
water.
• Dry with clean cloth or towel.
To clean the cable:
• Use a cloth or towel only lightly moistened with cleaner/ disinfectant (as recommended in TEE
Probe Care Card).
• Avoid too much moisture on the cloth/ towel as this may cause cleaning fluids to flow along the cable
and into the connector/ handle.
• Always clean the cable from the connector towards the middle of the cable.
• Hold the connector higher than the cable to avoid that cleaning fluids flows along the cable and into
the connector.
• Use the same technique when cleaning the cable from the handle side.
• To remove residual chemicals follow the same prosedure as above but use a cloth or towel lightly
moistened with clean water.
• Wipe off moisture with a cloth or towel and place the probe in a TEE Storage Rack and let it dry for
30 min.
Figure 10-1 Cleaning cable