10 - 16 Section 10-5 - System maintenance
10-5-6-5 Cleaning TEE probes
To clean the handle:
Use a cloth or towel only lightly moistened with cleaner/ disinfectant (as recommended in TEE
Probe Care Card).
Avoid too much moisture on the cloth/ towel.
To remove residual chemicals wipe the handle with a cloth or towel lightly moistened with clean
water.
Dry with clean cloth or towel.
To clean the cable:
Use a cloth or towel only lightly moistened with cleaner/ disinfectant (as recommended in TEE
Probe Care Card).
Avoid too much moisture on the cloth/ towel as this may cause cleaning fluids to flow along the cable
and into the connector/ handle.
Always clean the cable from the connector towards the middle of the cable.
Hold the connector higher than the cable to avoid that cleaning fluids flows along the cable and into
the connector.
Use the same technique when cleaning the cable from the handle side.
To remove residual chemicals follow the same prosedure as above but use a cloth or towel lightly
moistened with clean water.
Wipe off moisture with a cloth or towel and place the probe in a TEE Storage Rack and let it dry for
30 min.
Figure 10-1 Cleaning cable