5-70 Section 5-17 - LINUX - Voluson Maintenance System
Operation:
1.) Select [Restore] in the displayed “Media Menu” (see: Figure 5-38 above).
NOTE: If the message: “Recovery system (Linux) needs to be updated ...” appears, confirm with YES
.
NOTE: If the currently installed software has to be upgraded to a newer version, a new software specific
“Permanent key” is required.
a.) Enter the appropriate “ Permanent Key” ( calculated in OKOS; http://3.187.187.9/OKOS),
select OK and confirm with Enter.
.If the entered Permanent key is correct, following window appears, confirm with Enter.
2.) Only if you are sure you have saved the “Image Archive” and “Full System Configuration”, select
Yes in the following dialog box and confirm with Enter.
NOTE: If the message: “Found obsolete image files on R:....” appears, select Yes and confirm with Enter.
Figure 5-39 enter permanent key
Figure 5-40 confirm key
Figure 5-41 confirm dialog with YES