GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
4-6 Section 4-3 - General Procedure
4-3-3 Sleep Mode
Sleep Mode can reduce the time to start up the system with a portable exam. It takes less than 30
seconds to start up the system from sleep state.
4-3-3-1 Activate Sleep Mode
1.) If the system is not in read mode, freeze the image.
2.) Press the ON/OFF button on the control panel and select Sleep.
3.) After the monitor turns to black and On/Off switch turns to amber color, unplug the power cable from
the receptacle.
4.) To wake up the system, press the On/Off switch again.
NOTE: System can boot from sleep both connected and disconnected from AC power.
NOTE: Sleep mode is available in battery mode and AC power mode. Once the battery runs out the system will
not operate. To turn on the system, the user needs to connect the AC power. The sustaining time of
sleep mode will be reduced due to the battery capacity being decreased by the number of usage.
Figure 4-5 Selecting Sleep
Sleep mode is available when B mode button and the probe selection menu are available.