GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
4-22 Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-6 Archiving Images
4-5-6-1 Save Image Archive
NOTE: A backup of the Image Archive always contains the Patient Archive (database containing patient
demographic data and measurements) + images of the selected exams.
1.) Press the F2
key on the control panel.
2.) In the “Utilities” menu select SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select BACKUP
and then click the IMAGE ARCHIVE tab.
4.) Click the SAVE
button.
CAUTION
!! CAUTION:
It is recommended to Save Image Archive (Section 4-5-6-1 on page 4-22) and Backup the Full
System Configuration (Section 4-5-3 on page 4-17) once a week.
Figure 4-24 System Setup - Backup - IMAGE ARCHIVE page
Figure 4-25 Image Archive Save - preparations
[Save] button
[Load] button
click [Next] button
[Advanced] button
choose archiving
span time
check mark if desired