EasyManua.ls Logo

GE Voluson P6 - 10-8-2 GEHC Leakage Current Limits for Voluson P8;VolusonP6

GE Voluson P6
620 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-15
10-8-2 GEHC Leakage Current Limits for Voluson™ P8/Voluson™P6
The following limits are summarized for IEC60601-1 standard including national deviations and IEC
62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical
equipment.
Measurement limits per IEC60601-1 standard including national deviations.
Table 10-17 Type CF Patient Leakage Limits - ECG leads
Table 10-14 Enclosure Leakage Current Limits -Accessible Surfaces not protectively earthed
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.1 mA* 0.1 mA***
Close Protective earth and close neutral with reverse polarity 0.1 mA* 0.1 mA***
Open Protective earth and close neutral with normal polarity 0.3 mA** 0.5 mA***
Open Protective earth and close neutral with reverse polarity 0.3 mA** 0.5 mA***
Open neutral and close Protective earth with normal polarity 0.5 mA* 0.5 mA***
Open neutral and close Protective earth with reverse polarity 0.5 mA* 0.5 mA***
Table 10-15 Earth Leakage Current Limits
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.3 mA** 0.5 mA***
Close Protective earth and close neutral with reverse polarity 0.3 mA** 0.5 mA***
Open neutral and close Protective earth with normal polarity 1 mA* 1 mA***
Open neutral and close Protective earth with reverse polarity 1 mA* 1 mA***
Table 10-16 Type BF Patient Leakage Limits - Non-Conductive Surface
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.1 mA* 0.1 mA***
Close Protective earth and close neutral with reverse polarity 0.1 mA* 0.1 mA***
Open Protective earth and close neutral with normal polarity 0.5 mA* 0.5 mA***
Open Protective earth and close neutral with reverse polarity 0.5 mA* 0.5 mA***
Open neutral and close Protective earth with normal polarity 0.5 mA* 0.5 mA***
Open neutral and close Protective earth with reverse polarity 0.5 mA* 0.5 mA***
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.01 mA* 0.01 mA***

Table of Contents

Other manuals for GE Voluson P6

Related product manuals