GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
10-16 Section 10-8 - Electrical Safety Tests
Close Protective earth and close neutral with reverse polarity 0.01 mA* 0.01 mA***
Open Protective earth and close neutral with normal polarity 0.05 mA* 0.05 mA***
Open Protective earth and close neutral with reverse polarity 0.05 mA* 0.05 mA***
Open neutral and close Protective earth with normal polarity 0.05 mA* 0.05 mA***
Open neutral and close Protective earth with reverse polarity 0.05 mA* 0.05 mA***
Conditions
Country
USA Others