Section 8-53
Replacement of FAN with Cable ASSY II
8-53-1 Manpower
One person, 30 minutes.
8-53-2 Tools
Philips screwdriver, Allen Wrench
8-53-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
2.) Pull out the NEST as described in Section 8-28 "Replacement of the NEST" on page 8-93.
8-53-4 Removal Procedure
1.) Unscrew 4 screws and remove fan connector.
8-53-5 Installation Procedure
1.) Install the new parts in the reverse order of removal.
NOTE: In case of installing FAN with Cable ASSY (GPN:5669586), replace screws to SCREW PH M4X30
(GPN:5176333) in FRU.
2.) Perform: FRU8-47: Replacement of FAN with Cable ASSY II - Functional Tests