GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-175
Section 8-57
Replacement of ECG Kit assy
8-57-1 Manpower
One person, 10 minutes.
8-57-2 Tools
Philips screwdriver
8-57-3 Preparations
Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
8-57-4 Removal Procedure
1.) Remove 5 probe holders.
2.) Unscrew 9 numbers of screws below the keyboard base.
Figure 8-251 Removing 5 Probe holders and unscrewing 9 screws below the keyboard base