GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
5-76 Section 5-17 - LINUX - Voluson Maintenance System
5-17-5-3-2 New Hard Disk Setup
NOTE: If not already done, change BIOS “Boot Priority Order” as described in Section 5-17-5-3-1 on page 5-74.
1.) Wait for automatic boot into the “Voluson Maintenance System”.
2.) In the displayed menus select [Media] -> [NewDisk] and then press ENTER
.
3.) The following warning message will be displayed.
4.) ONLY if you are absolutely sure, select YES
and then confirm with ENTER.
5.) Creating a new disk will take some time......
Figure 5-47 Media -> NewDisk
Figure 5-48 WARNING! WARNING! WARNING!
NOTICE
If you select YES, all data on the entire hard disk will be lost and can not be recovered!
This includes all stored images and patient data!