GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-7
Table 7-1 list of shortcuts + description of what they do
Shortcut Description Access Level
Ctrl + 3
Factory use ONLY! This function is not intended for use in the field.
dongle
Ctrl + A
opens connected probe status dialog
dongle
Ctrl + B
yello US Image border display on/off
dongle
Ctrl + C
copy text
user
Ctrl + Alt + C
shows console test dialog
dongle
Alt + D
export system state (Full Backup and Dump files) to D:\export
user
Ctrl + Alt + D
Switch DICOM SR Follicle Format
user
Ctrl + Alt + E
event manager watch shows a list of executed events in dialog
dongle
Ctrl + H
shortcuts list (see Figure 7-5 above)
user
Ctrl + Alt + I
Start/stop presentation mode
user
Alt + L
opens log configuration dialog
dongle
Ctrl + M
mode control on/off
dongle
Alt + M
displays mouse catch (allows movement of mouse cursor on complete screen)
dongle
Alt + N
displays Network sniffer window (seeSection 7-4-2 "Capturing Network Logs with
Network Sniffer" on page 7-10)
dongle
Ctrl + P
change probe (dialog for “dummy” probes)
dongle
Alt + P
stores screenshot on D:\export
user
Shift + Ctrl + P
Change Probe Backward
dongle
Ctrl + S
User Program save
user
Alt + S
displays the hardkey (console) simulator (Figure 7-31), very important for remote access
(see7-5-5 "Virtual Network Computing (VNC)" on page 7-26)
dongle
Shift + Alt + S
opens the system setup dialog
dongle