GE RAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018)  BASIC SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
Endocavity Probe 
Leakage 
Current Checks
•
Also after corrective 
maintenance or as required by 
your facilities QA program.
Measurement Accuracy 
Checks
•
Also after corrective 
maintenance or as required by 
your facilities QA program.
Probe/
Phantom(optional) 
Checks
Check axial and lateral resolution (see Basic 
User Manual Technical specifications). 
Check Gain and TGC changes, vary the 
focus and check reaction on screen. 
•
Also after corrective 
maintenance or as required by 
your facilities QA program.
Functional Checks of all 
probes Section 10-6-2 
on page 10-9
•
Also after corrective 
maintenance or as required by 
your facilities QA program.
Table 10-2   Customer Care Schedule
Item Service at Indicated Time D W M A Notes