The selected Thermal Index is displayed in the [TI select] key.
Remarks:
•
While you are scanning, notice the index numbers you are using and which controls
affect the readings.
•
Try to keep the index numbers as low as you can, while maintaining diagnostic
information within the image. This is particularly important when scanning the fetus.
13.1.5 Display of Biopsy Guideline
•
The biopsy lines must be programmed once by the service personnel or the user. The
procedure must be repeated if probes and/or biopsy guides are exchanged!
•
Before performing a biopsy make sure that the displayed biopsy line coincides with the
needle track (check in a bowl filled with approx. 47˚C warm water). For further
instructions review: To program a Single Angle Biopsy Line
(chapter
'To program a
Single Angle Biopsy Line'
on page 5-15
)
To program a Multi Angle Biopsy Line
(chapter
'To program a Multi Angle Biopsy Line'
on page 5-16
)
•
Please read the “Instructions for safe Use” in Probes and Biopsy /
Biopsy Special
Concerns
(chapter
'Biopsy Safety and Maintenance'
on page 2-19
)
.
Operation:
Select the [Utilities] and the [Biopsy Line] key. The Biopsy guideline appears on the monitor.
The Biopsy line is switched on if the [Biopsy Line] key is highlighted. To switch OFF the Biopsy
guideline select the [Biopsy Line] key again.
The [Biopsy Line] key toggles between ON and OFF.
•
To program a Biopsy guideline review: To program a Single Angle Biopsy Line
(chapter
'To program a Single Angle Biopsy Line'
on page 5-15
)
To program a Multi Angle Biopsy
Line
(chapter
'To program a Multi Angle Biopsy Line'
on page 5-16
)
Remarks:
•
For Handling, Sterilization, Mounting a biopsy guide, etc. review: Biopsy Special
Concerns
(chapter
'Biopsy Safety and Maintenance'
on page 2-19
)
.
•
For each probe one biopsy guideline is programmable.
13.1.6 Lock Screen
13.1.6.1 Introduction
Utilities and System Setup
13-8
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4