EasyManua.ls Logo

GE Voluson S6 - Power-Doppler Mode (PD Mode)

GE Voluson S6
684 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
V-P (velocity + power)
T (turbulence)
P-T (power + turbulence)
CF window size: axial: 0 to B scan range
lateral: 0 to B scan range
Baseline shift: 17 steps (independent from spectral Doppler)
Inversion of color direction: yes
Wall Motion Filter: 8 steps (low1, low2, mid1, mid2, high1, high2, max1, max2)
Smoothing Filter: 12 steps rising time
12 steps falling time
Gain control: +15dB to -15dB, 0.2dB steps
Line Density (color line
density):
10 steps
Ensemble CF: 7 to 31
(color shots per line): MCF: 8 to 16
Flow Resolution: 4 steps (low, mid1, mid2, high)
Pulse repetition frequency: CF: 150 Hz to 20.5 kHz
MCF: 150 Hz to 20.5 kHz
CF Map: 8 Color maps can be selected
Frequency range: 1 to 16 MHz depending on the probe, adjustable in 3 steps (low, mid, high)
Balance: from 25 to 225
Max. meas. velocity: 4.23 m/sec
Min. meas. velocity: 0.3 cm/sec
Scale: kHz, cm/s, m/s
Automatic moving tissue
suppression:
yes
16.10.5 Power-Doppler
Screen formats: 2D+PD (single, dual, quad)
Display modes: Simultaneous dual mode: 2D/2D+PD
Triplex mode: 2D+PD/PW
Volume Mode: 3D+PD;
PD coding steps: 256 color steps
PD window size: lateral: maximum to minimum B mode scan angle
axial: B-scan range
Display mode: P (power)
Wall motion Filter: 8 steps (low1, low2, mid1, mid2, high1, high2, max1, max2)
Smoothing Filter: rising edge: 12 steps