9.8.2 After the VCI-A Acquisition
After [Freeze] the system switches automatically to freeze mode and the “Vol. Cine” menu
appears. The selected format and the last acquired volume will be present on the monitor.
Upon [Freeze] a certain number of volumes will be stored in the cine memory. The sequence
can be reviewed volume by volume.
'Volume Cine'
on page 9-95
9.9 VCI-Omniview
Note
VCI-OmniView is an option. If this option is not installed, the [VCI-OmniView] key is hidden.
Note
SingleView is an option for Voluson® S6 only. This will be activated when SingleView option is
installed and Advanced VCI (VCI-OmniView) option is not installed. SingleView supports only
basic controls of VCI-OmniView with Line method.
By setting the necessary sweep angle for the desired ROI, the system provides a coronal
plane (Omni View). The rendering box is very thin and so you can visualize the tissue
information of a thick slice. A mixture of surface texture and transparent maximum (or X-ray)
rendering modes (70/30) plus a low setting of surface transparency (20-50) is used. The
resulting image shows the average (integrated) gray value of the tissue contained within the
narrow box. Omni View improves the contrast resolution and the signal / noise ratio and
therefore facilitates the detection of diffuse lesions in organs. The result is an image with no
speckle pattern and a highly improved tissue contrast.
The “4D Mode” menu appears on the menu area (scan mode).
Volume Mode
9-102
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4