3. Start the Pre-use check by pressing Yes.
SVS-5073_XX
4. Follow the on-screen instructions.
5. After pressing the touchpad Pre-use
check the message Do you want to start
a pre-use check? is displayed - confirm
by pressing Yes.
6.2.2 Internal leakage test
SVS-0082_XX
Connect the test tube between the inspiratory
outlet and the expiratory inlet.
Important: Use only a test tube from the
manufacturer of the ventilator system.
6.2.3 Battery switch test
The Pre-use check tests the ventilator’s ability
to switch between AC and battery power when
AC power is lost and restored:
• When the on-screen instruction appears,
disconnect the ventilator from AC power.
• When the on-screen instruction appears,
reconnect the ventilator to AC power.
Note: This test will not be performed if there
is less than 10 minutes of battery operation
available.
106
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 6
| Operation overview |