11.8 Breathing parameters: default values & allowed settings (standard
configuration)
Setting rangeFactory set defaultParameter
-2Bias flow (l/min)
1 - 154Breath cycle time, SIMV (s)
4 - 10015CMV frequency (b/min)
ON/ OFFOFFCompensate for compliance
1 - 102Flow trig sensitivity level
1:10 - 4:11:2I:E ratio
1:10 - 4:11:2I:E ratio in backup
1 - 7030End inspiration (% of peak flow)
10 - 7050End inspiration (% of peak flow) in NIV
0 - 205Inspiratory rise time (%)
0 - 0.40.15Inspiratory rise time (s)
0 - 0.40.2Inspiratory rise time (s) in NIV
-3.3Maximum inspiratory flow (l/s)
-120Maximum permitted absolute pressure (cmH
2
O)
-62Maximum permitted absolute pressure in NIV
(cmH
2
O)
0.5 - 607. 5Minute Volume (l/min)
-VCMode (Invasive ventilation)
-PSMode (Non invasive ventilation)
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
153
| Technical data |
11 |