Setting rangeFactory set defaultParameter
21 - 10040O
2
concentration (%)
0 - 508PEEP (cmH
2
O)
2 - 205PEEP in NIV (cmH
2
O)
(PEEP+1) - 5015Phigh (cmH
2
O)
-20 - 0-Press trig sensitivity level (cmH
2
O)
0 - (120 - PEEP)15Pressure level above PEEP (cmH
2
O)
0 - (62 - PEEP)5Pressure level above PEEP in NIV (cmH
2
O)
5- (120 -PEEP)15Pressure level above PEEP in backup (cmH
2
O)
5 - (62 – PEEP)5Pressure level above PEEP in NIV backup (cmH
2
O)
0 - (120 - PEEP)0PS above PEEP in Bi-Vent/APRV (cmH
2
O)
0 - (120 - P
High
)0PS above Phigh in Bi-Vent/APRV (cmH
2
O)
4 - 10015Resp rate in backup
1 - 605SIMV frequency (b/min)
0.2 - 30 s2Thigh (s)
0.1 - 50.9Ti (s)
0.1 - 50.9Ti in backup (s)
100 - 2000500Tidal Volume (ml)
100 - 2000500Tidal volume in backup (ml)
0 - 3010Tpause (%)
0 - 1.50.4Tpause (s)
0.1 - 10 s2TPEEP (s)
10 - 25050Weight (kg)
154
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 11
| Technical data |