EasyManua.ls Logo

Getinge Maquet SERVO-s - Differences from Invasive Pressure Control Mode

Getinge Maquet SERVO-s
185 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.11.2 Functional description Pressure
Control
The NIV Pressure Controlled mode is a
controlled breathing mode.
SVS-9013_XX
The following parameters are set:
1. PC (Pressure Control level) above PEEP
(cmH
2
O)
2. Respiratory Rate (b/min)
3. PEEP (cmH
2
O)
4. Oxygen concentration (%)
5. I:E ratio / Insp. time
6. Inspiratory rise time (%/s)
Differences from invasive Pressure control
mode:
When the Standby key is pressed a waiting
position dialog is shown. All patient related
alarms are turned off during 120 seconds.
Press the Start ventilation touchpad to start
the ventilation.
During NIV the ventilator automatically
adapts to variations in leakage in order to
maintain the required pressure and PEEP
level. If leakage is excessive, the ventilator
will issue a high priority alarm and deliver a
flow according to settings. Ventilation will
resume automatically if the leakage
decreases. Ventilation can also be started
manually by pressing the Resume ventilation
touchpad in the dialog.
Trigger sensitivity cannot be set in NIV.
Detection of pressure below PEEP or
expiratory volume decrease will start a new
breath.
Read about alarm settings on page 128.
See also Set Ventilation Mode on page 114.
92
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 5
| Ventilation, modes and functions |

Table of Contents

Related product manuals