7.2.2 Responding to Alarms
To respond to a High or Medium priority alarm:
1. If desired, press the Audio Pause fixed
key for less than two seconds to silence
the alarm for two minutes.
2. Take action to resolve the alarm condition.
3. Press the Audio Pause key to reset the
latched high priority alarm and clear the
message from the screen.
The Audio Pause key is identified by the
symbol.
To respond to a Low priority alarm:
1. If desired, press the Audio Pause key for
less than two seconds to reset the alarm
even if the alarm condition remains.
2. Take action to resolve the alarm condition.
The alarm is automatically reset once the
alarm condition ceases.
Note:
The following alarms display an Audio off?
message when activated:
Air Supply Pressure: Low
O
2
Supply Pressure: Low
Battery Operation
For these alarms, you can silence the audio
signal even if the alarm condition is not
resolved. However, the system will
eventually reactivate the alarm.
Using the Audio Pause Key
SVX-5098_EN
Pressing the Audio Pause fixed key for less
than two seconds has the following results:
Active alarms are silenced for two minutes.
A crossed bell symbol along with the time
remaining in the silent period are displayed
in the message area.
Each press of the Audio Pause key restarts
the two minute silent period from when the
user last pressed the key.
Latched alarms are reset if the alarm
condition has ceased.
Note: The No battery capacity alarm and
technical alarms cannot be silenced.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
125