6.7 Set alarm limits
SVS-5087_XX
To set alarm limits:
1. Press the Alarm Profile fixed key.
2. Press the touchpad corresponding to the
alarm limit you want to adjust or press the
Alarm sound level touchpad.
3. Turn the Main Rotary Dial to adjust values.
4. Confirm each setting by pressing the
parameter touchpad or Main Rotary Dial.
5. Press Autoset, if desired, to get a proposal
for alarm limits in VC, PC, and PRVC
modes.
Important: Before accepting Autoset
values, make sure they are appropriate
for the patient. If not, enter settings
manually.
6. Press Accept to activate the new alarm
limits.
Important: Before accepting Autoset
values, make sure they are appropriate for
the patient. If not, enter settings manually.
Notes:
• Autoset is not possible in Standby
because the ventilator requires patient
values in order to propose alarm limits.
• Autoset is not available in supported or
NIV (optional) modes.
• Current alarm limits are displayed during
ventilation in smaller figures to the right
of the parameter display.
• The Main Rotary Dial becomes
inoperative for 2 seconds when a defined
safety limit for the alarm limit being
adjusted is reached. After this, it will be
operable again.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
115
| Operation overview |
6 |