Airway Pressure
Range: -40 - 160 cmH
2
O
Inaccuracy: ±(1 cmH
2
O + 5 % of actual
value)
Supply Pressure
Range: 0 - 7 bar
Inaccuracy: ± 5% of read value
Filtering
Pressure waveform: Low pass filtered (time
constant 15 ms)
The measured and calculated values displayed
or used for control have in some cases been
subjected to filtering and smoothing
techniques. This is done to capture the
important patterns in the data while excluding
noise and make the data shown clinically
relevant. These techniques are part of the
inaccuracy specified in the technical data.
152
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual