No consistent patient effort
No consistent patient effort is an alarm in
invasive ventilation.
If the patient fulfils the criteria for the No
consistent patient effort alarm, the ventilator
will lock in backup ventilation.
A dialog You are in Backup ventilation. Review
ventilation settings or continue in support
mode is displayed on the screen. As long as
this dialog is open no other menu window or
dialog can be accessed. Only the Standby
touchpad is active.
The user will be given the following choices:
1 2
1. Review vent. settings
2. Continue in support mode
The ventilator will remain in backup ventilation
until a choice is made.
Review ventilation settings
Press Review vent. settings in the Backup
ventilation window to return to the Set
Ventilation Mode window.
3
21
The following functions are available:
1. Previous Mode - recalls the previous
accepted ventilation mode.
2. Cancel - closes the Set Ventilation Mode
window without changes being applied,
i.e. ventilation will continue as before.
3. Accept - accepts the settings and
continues in the supported mode with
reset apnea time.
Continue in support mode
Press Continue in support mode in the Backup
ventilation window to continue in support
mode. The apnea time will be reset.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
95
| Ventilation, modes and functions |
5 |