11.3 Standard condition specification
Inaccuracy ranges in this document assume
the following standard conditions and the
worst case, i.e. all errors are summarized
positive. Statistically 95% of all values will be
within 2/3 of the given error.
Ambient pressure: 101.3 kPa
Room temperature: 20 ºC
Inlet pressure: 4.3 kPa x 100
Pre-use check performed on a warmed up
ventilator
Pre-use check performed with 99 %
oxygen content in O
2
supply.
Pre-use check performed with 21 ±0.5%
oxygen content in air supply.
Patient circuit compliance compensation is
activated.
Default settings unless otherwise specified
All measured, preset and indicated flows
and volumes are referenced to BTPS, see
chapter Definitions on page 165.
All measured, inlet gas pressures and flows
are referenced to STPD, see chapter
Definitions on page 165.
Humidifier water level is between 50 % and
100 % of the maximum level.
Set I:E is less than 1:1.
Set ventilatory frequency is less than or
equal to 100 breaths/minute.
Inaccuracy does not include measurement
uncertainty. Values including measurement
uncertainty are specified in the service
documentation available from the
manufacturer.
Constant leakage fraction below 30 % in
NIV modes.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
149
| Technical data |
11 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-s and is the answer not in the manual?

Getinge Maquet SERVO-s Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-s
CategoryFan
LanguageEnglish

Related product manuals