Direct Access Knobs - Safety
The four Direct Access Knob parameters are
displayed at the bottom of the screen with
color-coded bars that indicate whether the
parameter values are within
generally-recognized safety limits.
SVS-5003_XX
The figure above shows the following
components.
1. A Direct Access Knob
2. A yellow bar indicating the corresponding
parameter value is outside safety limits; a
text message is displayed accompanied
by an audible signal.
3. A red bar indicating the corresponding
parameter value is significantly outside
safety limits; a text message is displayed
accompanied by an audible signal.
Note: When the defined safety limits for a
given parameter have been reached, the
Direct Access Knob becomes inoperative
for 2 seconds to indicate that a safety limit
has been reached.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
25
| System Overview |
2 |