6.2.5 Compensate for Circuit Compliance
When the Compensate for circuit compliance?
dialog appears on the screen, do one of the
following:
To add the compensation, press Yes
(recommended),
To refuse the compensation, press No.
Important: If the patient circuit is changed,
a Patient Circuit Test must be performed.
Note: Circuit compliance compensation is
not available in NIV modes (option).
6.2.6 Test Alarm Output Connection
(Option)
If the Alarm Output Connection option is
installed, a dialog for the external alarm system
test appears on the screen.
Do one of the following:
To perform the test, press Yes and follow
the on-screen instructions.
To cancel the test, press No.
108
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual