11.12 Log function
Event Log
• Alarms
• Ventilator settings
• Apnea periods
• Immediate functions
It is possible to display up to 1000 recorded
events in chronological order.
Service Log
• Technical alarms
• Test results
• Preventive maintenance
• Service report history
• Configuration log
11.13 Immediate functions
Oxygen Breaths
100% for 1 minute.
Start Breath
Initiation of 1 breath in all modes.
(In SIMV mode initiation of one mandatory
breath).
Pause Hold
Inspiratory or expiratory.
11.14 Communication/interface
Serial Port
RS-232C-isolated. For data communication
via the Communication Interface Emulator
(CIE).
Information regarding connector wiring and
EMC performance is available from the
manufacturer.
Communication Interface Emulator (CIE)
A protocol for data communication with
external devices. The SERVO-s Ventilator
System software v8.0 uses CIE Protocol
version 004. This protocol is compatible with
previous versions although all data may not
be available.
Alarm Output Connection (Option)
Isolated 4-pole modular connector for
communication of high and medium priority
alarms. The alarm output connection option
is a non-guaranteed alarm in accordance with
IEC60601-1-8.
Switching capability: Max 40 V DC, Max 500
mA, Max 20 W.
Information regarding connector wiring and
EMC performance is available from the
manufacturer.
158
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual