9.4 Low priority alarms
RemediesPossible causesAlarm message
Check the connection to AC power.AC power interrupted.Battery operation
Check screen and knobs.
Contact a service technician.
Screen or knob has been pressed for
more than one minute. Screen or knob
hardware time out.
Touch screen or
knob press time
exceeded
9.5 Technical error messages
RemediesCausesError code number
Restart the ventilator and perform a
Pre-use check.
Contact a service technician.
Technical problem, identified by the error
code xxxx.
xxxx (General)
Contact a service technician.Internal power error.1 - 6, 29, 10001
Contact a service technician.Control system error7, 10-12, 16
Contact a service technician.Battery information error43
Check that the loudspeaker outlet is not
obstructed. Restart the ventilator and
perform a Pre-use check.
Contact a service technician.
Audible alarm/ loudspeaker error28, 20004
Contact a service technician.Internal clock error41
Contact a service technician.Internal communication error25, 33-35, 50,
10002, 20001
Contact a service technician.Barometer error38-39
Contact a service technician.Alarm output connection error46
Contact a service technician.Fixed battery missing47
Contact a service technician.Timeout error8-9, 48-49
Restart the ventilator and perform a
Pre-use check.
Contact a service technician.
Backup audible alarm error22, 24, 27
Contact a service technician.Backlight error20002
Check the user interface buttons.
Contact a service technician.
Membrane button error20003
Contact a service technician.Technical problem with On/Off switch.51
Contact a service technician.Exp. flow meter error40001 - 40011
Restart the ventilator and perform a
Pre-use check.
Contact a service technician.
Other error40, 42, 44, 45, 50,
54, 10003, 20005
140
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 9
| System messages |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-s and is the answer not in the manual?

Getinge Maquet SERVO-s Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-s
CategoryFan
LanguageEnglish

Related product manuals