3.3.2 Status Messages
ExplanationMessage (message type)
Power failure.Technical error no. 1 - 6, 29, 10001 (technical error)
There is a problem with the battery modules. One
or more battery modules must be replaced.
Check battery status (Text Message)
AC power is off line due to a power failure or
disconnection.
Battery operation (Low Priority Alarm)
Less than 10 minutes left of battery operation.Limited battery capacity (Medium Priority Alarm)
Less than 3 minutes left of battery operation.No battery capacity (High Priority Alarm)
Battery voltage too low. Cannot guarantee
continued ventilator operation.
Low battery voltage (High Priority Alarm)
3.3.3 AC Power Failure
In the event of an AC power failure or
disconnection, the ventilator switches to
battery operation and activates low/medium
priority alarms, see Table above in Status
Messages for details.