7.2.3 Pre-silencing Alarms
To silence most alarms (active and inactive)
for two minutes, press and hold the Audio
Pause fixed key for more than two seconds.
This action has the following results:
1. All alarms, active and inactive, are
silenced for two minutes from the time the
key was pressed.
2. A double crossed bell symbol along with
the time remaining in the silent period are
displayed in the message area.
3. Pressing the Audio Pause key again for
less than two seconds will now restart the
silent period for two additional minutes
from when the user last pressed the key.
4. Latched alarms are reset if the alarm
condition has ceased.
SVX-5098_EN
When the alarms are pre-silenced, pressing
and holding the Audio Pause key again for
more than 2 seconds will reactivate all audible
alarms.
126
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 7
| Alarms |