10.1 Introduction
This section provides instructions for
accessing and editing the start-up
configuration and changing the type of
ventilation.
The ventilator will always start up with the
stored Start-Up Configuration. The Start-Up
Configuration can be edited, copied, and
saved.
You can edit the following Start-Up
Configuration settings:
• Type of ventilation - invasive or non invasive
(NIV)
• Volume setting
• Breath cycle setting
• Pre/post oxygenation concentration above
set O
2
concentration (%)
• Option to deactivate backup ventilation
(Disabled or Enabled)
• NIV disconnect function (Low flow, High
flow or Disabled)
• Extended leakage test during pre-use check
• Volume Control with alternative flow
patterns
10.2 Access the Start-Up
Configuration
SVS-9046
Note: The ventilator must be in Standby.
1. Press the Menu fixed key.
2. Press the Biomed touchpad and enter the
access code (the factory setting is 1973).
The Biomed submenu consists of the following
touchpads:
• Service
• Edit configuration
• Copy configuration
• Set date and clock
• Change access code
• Start MCare Remote Service
To alter ventilator settings, press the
appropriate touchpad and follow on-screen
instructions.
142
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 10
| Start-up configuration |