6.2.7 Complete the Pre-use check
A message appears on screen for each
Pre-use check test, as appropriate: Cancelled,
Failed, Not Completed, Passed or Running.
Press OK to confirm and to have the Pre-use
check tests logged. The ventilator now
switches to Standby.
Notes:
After the Pre-use check is completed (or
skipped), you will be prompted to keep
or discard old patient-related data.
By accessing the Status menu, the
results of the two latest Pre-use checks
are displayed under General.
The status of the Patient Circuit test is
displayed under Status/ Patient Circuit.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual