13.7 Screen touchpads
When the touchpad for a selected ventilation mode is pressed, a window
appears with the valid settings for this mode. The window has an Accept
touchpad and a Cancel touchpad. When the Accept touchpad is pressed the
ventilator starts to ventilate with the new settings. If the Cancel touchpad is
pressed this window will disappear and the ventilator will continue to ventilate
with the original settings. To support the clinician in adjusting settings some
values, derived from settings, are shown in the upper right field of the set
ventilation mode window e.g. inspiration time in seconds, calculated inspiratory
flow.
Mode Xxxxx
When the admit patient function is activated the clinician can enter or amend
patient details:
date of admissionpatient name
••identity number height
weightdate of birth
Admit patient
In the Status function, an icon displays which power source is currently active
i.e. Mains, Ext. 12 V, or battery. When the batteries are in use, information
about remaining time is also displayed. When the Status touchpad is activated
a windows appears displaying the status of:
Status of batteriesGeneral system information
••Status of O
2
cell/sensor Installed options
Status of Pre-use check
Status of Expiratory cassette
Status
General
O
2
cell/sensor
Exp. cassette
Batteries
Installed options
Pre-use check
176
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 13
| Appendix • User interface |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-s and is the answer not in the manual?

Getinge Maquet SERVO-s Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-s
CategoryFan
LanguageEnglish

Related product manuals