The Check tubing alarm is turned off: the
maximum duration of the preparation phase
is 120 seconds. After 120 seconds, the system
automatically returns to ventilation using the
previous oxygen setting.
Note: The Cancel touchpad will close the
Suction Support program.
6.10.3 Disconnect Phase
The system automatically enters the
disconnect phase when the patient is
disconnected during the
pre-oxygenation/preparation phase.
During the disconnect phase the following
alarms are turned off for up to 60 seconds:
• Apnea
• Minute volume
• Respiratory rate
• PEEP
When the patient is reconnected, the system
automatically enters the post-oxygenation
phase and restarts ventilation.
It is also possible to restart ventilation
manually by pressing the Start ventilation
touchpad.
6.10.4 Post-Oxygenation Phase
After reconnection, the ventilator will deliver
the same oxygen concentration as in the
preparation phase for 60 seconds.
After 60 seconds the system automatically
returns to ventilation using the previous
oxygen setting.
6.10.5 Closed-suction systems
If a closed-suction system is used, the Suction
Support, Insp. hold and Exp. hold functions
should not be used. Pressure-based modes
(such as Pressure Control, Pressure Support,
Bi-Vent/APRV, or SIMV (PC) + PS) should be
used. Settings should be adjusted to levels
suitable for the patient. Hospital guidelines for
suctioning should be followed.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
119