1.2.6 Electromagnetic Compatibility
Important: The ventilator system must be
installed and put into service according to
Electromagnetic Compatibility, SERVO-s
Ventilator System.
In order to ensure that the SERVO-s Ventilator
System, during electromagnetic disturbances,
will deliver ventilation at the connection port
within the alarm limits set by the user, or
generate an alarm condition, the following
essential performance (IEC 60601-1) has been
monitored during electromagnetic immunity
tests:
Delivered volume
Monitoring of:
Oxygen concentration-
- Airway pressure
- Expired volume
- Electrical supply failure
- Internal electrical power source
- Gas supply
Ability to generate alarms
No degradation or failure of the essential
performance has been observed during these
tests.
No other effects on the ventilator system have
been observed during the electromagnetic
immunity tests.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
13