EasyManuals Logo
Home>Getinge>Fan>Maquet SERVO-s

Getinge Maquet SERVO-s User Manual

Getinge Maquet SERVO-s
185 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
CAUTIONS:
It is not recommended to use a nebulizer
during NIV as the nebulized drug might
come in contact with the patient eyes in
case of leakage.
Mask leakage might affect the efficiency
of a nebulizer.
Important:
The mask must be applied correctly in
order to avoid leakage.
Selection of a mask must take into
consideration its proper size and
accurate adaptation to the patient.
CO
2
rebreathing will increase during NIV
and use of a face mask.
Read about alarm settings on page 128.
See also Set Ventilation Mode on page 114.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
91
| Ventilation, modes and functions |
5 |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet SERVO-s and is the answer not in the manual?

Getinge Maquet SERVO-s Specifications

General IconGeneral
BrandGetinge
ModelMaquet SERVO-s
CategoryFan
LanguageEnglish

Related product manuals