CAUTIONS:
It is not recommended to use a nebulizer
during NIV as the nebulized drug might
come in contact with the patient eyes in
case of leakage.
Mask leakage might affect the efficiency
of a nebulizer.
Important:
The mask must be applied correctly in
order to avoid leakage.
Selection of a mask must take into
consideration its proper size and
accurate adaptation to the patient.
CO
2
rebreathing will increase during NIV
and use of a face mask.
Read about alarm settings on page 128.
See also Set Ventilation Mode on page 114.
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual