13.1 Fixed keys
Audio pause (Silence or pre-silence alarms)
By pressing Autoset in controlled modes of
ventilation the alarm limits are automatically set for:
• Pressure
• Volume
• Resp. Rate
• PEEP
In spontaneous modes an alarm setting for apnea
time is available.
Note: Autoset is not possible in NIV.
Alarm profile setup. Possible
selections:
• Pressure (upper)
• Minute volume (lower and
upper)
• Respiratory rate (lower and
upper)
• End expiratory pressure (lower
and upper)
• Alarm sound level (10-100%)
Note: In NIV the alarm sound
can be permanently silenced
(audio off).
Alarm profile
The trend graph appears when the Trend key is pressed. Data can be recorded over
a period of time up to 24 hours. The time resolution is displayed in the trend graph.
Trends
168
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 13
| Appendix • User interface |