13.1 Fixed keys
Audio pause (Silence or pre-silence alarms)
By pressing Autoset in controlled modes of
ventilation the alarm limits are automatically set for:
Pressure
Volume
Resp. Rate
PEEP
In spontaneous modes an alarm setting for apnea
time is available.
Note: Autoset is not possible in NIV.
Alarm profile setup. Possible
selections:
Pressure (upper)
Minute volume (lower and
upper)
Respiratory rate (lower and
upper)
End expiratory pressure (lower
and upper)
Alarm sound level (10-100%)
Note: In NIV the alarm sound
can be permanently silenced
(audio off).
Alarm profile
The trend graph appears when the Trend key is pressed. Data can be recorded over
a period of time up to 24 hours. The time resolution is displayed in the trend graph.
Trends
168
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual
| 13
| Appendix • User interface |