13.5 Menu Key
Alarm profile setupAlarm
By pressing Autoset in controlled
modes of ventilation the alarm limits
are automatically set for:
• Pressure
• Volume
• Resp. Rate
• PEEP
In spontaneous modes an alarm
setting for apnea time is available.
Note: Autoset is not possible in
NIV.
Alarm profile setup. Possible
selections:
• Pressure (upper)
• Minute volume (lower and upper)
• Respiratory rate (lower and upper)
• End expiratory pressure (lower and
upper)
• Alarm sound level (10-100%)
Note: In NIV the alarm sound can
be permanently silenced (audio off).
Alarm:Profile
This shows alarms that have been activated. The list is in alphabetical order.Alarm:History
Audio pause (Silence or pre-silence alarms)
Alarm:
Review
Review trends, event log or configuration.Review:Trends
Review:Event log
Review:View
configuration
172
SERVO-s VENTILATOR SYSTEM V8.0, User's Manual