GDS-2000 Series User Manual 
160 
 
3.  Press F5 (More) twice. 
More
F 5
 
x2 
 
4.  Press F1 (ProbeComp 
Menu). 
ProbeComp
Menu
F 1
 
5.  Press F1 (Wavetype) 
repeatedly to select the 
standard square wave. 
 
 
6.  Press the Auto Set key. The 
compensation signal 
appears on the display. 
Auto Set
 
 
7.  Press the Display key, then 
F1 (Type) twice to select the 
vector waveform. 
Display
Type
Vectors
F 1
 
 
 
8.  Turn the adjustment point on the probe until 
the signal edge becomes sharp. 
 
Under
Compensation
Normal
Over
Compensation