MAINTENANCE
159
8. The calibration for Channel1
starts and ends
automatically, in less than 5
minutes.
Ch1 calibration 1/3
9. When finished, connect the
calibration signal to
Channel2 and press F5.
Channel2 calibration starts.
Done!!
10. (for 4 Channel model only) Repeat the above
step for Channel 3 and 4.
11. When the calibration for all channels is
completed, the display goes back the default
state.
Probe Compensation
Panel operation
1. Connect the probe between Channel1 input and
the probe compensation output (2Vp-p, 1kHz
square wave) on the front panel. Set the probe
attenuation to x10.