38
Trigger key (2/5)
Type
Pulse
F 1
Source
CH1
F 2
Mode
Auto
F 3
Slope /
Coupling
F 5
TRIGGER
Auto/ Normal/
Single
When <
20.0ns
F 4
To Slope/Coupling
>/ </ =/ =
20ns~200us
(4CH) CH1/2/3/4/Line
(2CH) CH1/2/Ext/Line
MENU
Edge/Pulse
Select Edge/Pulse trigger type
F1
Select trigger source
F2
Select trigger mode
F3
Select pulse trigger condition
and pulse width
F4
VAR
Go to slope/coupling menu
F5
Trigger key (3/5)
100ns~1.3ms
TTL: 1.48V/
ECL: -1.35V
User: -12~+12V
2 ~ 65000
Type
Delay
F 1
By Time
100ns
F 2
By Event
2
F 3
Slope/
Coupling
F 5
TRIGGER
Ext:
TTL
F 4
(2CH Only)
To Slope/Coupling
MENU
Select Delay trigger type
F1
Select time delay mode and delay
length
F2VAR
Select event delay mode and
event count
F3VAR
Select external trigger type and
adjust trigger level (User type)
F4VAR
Go to slope/coupling menu
F5