Chapter 6 Maintenance
Honeywell Confidential: Use or disclosure of information on this page is subject to the restrictions on the title page.
13841 Rev A
210
6.3.1.4
Removal
Remove batteries from the Vibrex unit, as follows:
1. Using a Phillips screwdriver, turn the large Phillips
screw on the unit’s battery end plate (15), one quarter
turn counter-clockwise.
2. Carefully remove the battery end plate.
3. Hold the unit with its connector subpanel pointing
upward.
4. Slide the batteries out of the side hand grips.
NOTE: The user may find out whether batteries are low on
power by using the Status feature. To access, press the
STATUS button from the Main Menu screen.
6.3.1.5
Reinstallation
When replacing the batteries it is important to observe the proper
orientation as depicted by the decals located inside the Vibrex
2000 battery tubes. Reinstall the batteries, as follows:
1. Insert the right-side batteries (in the hand grip next to the
contrast buttons) negative end first. The raised contact
must be facing out.
1. 2. Install the left-side batteries (in the hand grip next to
the context keys) positive end first.
2. Reinstall the battery end plate (15) in the reverse order
of removal.
3. For an illustration, see the diagram in Section 6.3.1.3
Battery Insertion Diagram. Diagrams on the inside of the
hand grips illustrate the correct battery insertion and
polarity.
NOTE: Failure to properly install the batteries could result
in a battery leaching its electrolyte onto the electronic
circuitry inside the Vibrex 2000, destroying the circuit
board. This is especially true if only one battery is reversed.