Model 5340A 
Maintenance and Service 
Figure 5
-
2.  Adjustment Procedures (Continued) 
8. 
QUAD DETECTOR ADJUSTMENT 
a.  Set 5340A RANGE switch to 
10 
Hz to 
18 
GHz and the RESOLUTION switch to 
100 
Hz. 
b.  Connect 608 Signal Generator to 50R Type N INPUT.  Set generator to 280 
MHz 
at 
-
20 
dBm. 
c.  Set A13R31 fully counterclockwise. 
d.  Measure voltage 
at 
XA4(8) (QUAD DET) and set A13R31 for 
-
450 mV 
f 
10 
mV. 
e.  Disconnect 280 MHz signal and connect 20 GHz 
at 
0 
dBm and observe 
a 
stable display.  If the display 
is 
unstable, adjust A7R27 (GAIN) counterclockwise for 
a 
stable display. If 
this 
adjustment is made, it 
is 
necessary to readjust the input loop.  Repeat step 
7. 
9. 
OSCILLATOR ADJUSTMENT 
- 
STANDARD OSCILLATOR 
a. 
Connect 
53 
OSC 
output, available on the 5340A rear panel to the input of 
a 
high resolution frequency 
counter such 
as 
a 
5360A.  Connect 
a 
suitable external frequency standard (such 
as 
HP 5061 Cesium Beam) 
to 
the external oscillator input of the 5360A. 
b.  Remove the A18 05340
-
60036 Oscillator board from the 5340A and note the frequency marked on 
the 
label.  Reinstall the board. 
Remove the screw covering the frequency adjustment. 
Use an insulated tuning tool to adjust the oscillator for 
a 
5360A display of the frequency noted in step b. 
c. 
d. 
10. 
OSCILLATOR ADJUSTMENT 
- 
OPTION 
001 
NOTE 
The 5340A must have primary power applied for 
at 
least 24 hours to allow the 
oscillator temperature to stabilize. 
a. 
Set 
controls on oscilloscope 
as 
follows: 
CHANNEL A 0.2 
V/cm; DC coupled 
+ 
up 
SWEEP 0.05 
psec/cm 
TRIGGER; EXT, 
+ 
slope, ACF 
MODE 
to NORM 
DISPLAY to CHANNEL A 
b.  Connect 
a 
suitable 
1 
MHz, 5 MHz or 
10 
MHz frequency standard (such 
as 
an HP Cesium Beam) to the 
EXT Input on oscilloscope. 
c. 
Connect Oscilloscope Channel A to 
10 
MHz output (53) available on 5340A rear panel. 
d.  Adjust oscillator FREQUENCY ADJ for minimum sideways movement of  the oscillator signal. 
5
-
10