EasyManua.ls Logo

HSD ES919 NC - Replacing and Calibrating the SC Sensor in Belt Drive Versions

Default Icon
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
______HSD Technological equipment for Automation _____________________________________
5801H0017 Rev. 02 _________________________________________________________ 89/98
13.3.4 Replacing and calibrating the SC sensor in belt drive versions
BEFORE STARTING ANY WORK ON THE ELECTRO-SPINDLE, READ AND
FOLLOW ALL THE SAFETY PRECAUTIONS AND MAINTENANCE SAFETY
WARNINGS.
SEE IN PARTICULAR CHAPTER § 5 AND PAGES 66 AND 71 OF CHAPTER § 12.
1.
Turn the C axis to any position other than the zero
reference position, i.e. to any position other than that
shown in Figure 13.4.
Disconnect the electro-spindle from the electrical supply.
Remove the eight fixing screws and remove the cover
from the body of the C axis unit (see figure on the right).
Disconnect the sensor's electrical connector, which is
located in the point shown by the arrow in the figure on
the right (see also position 5 of Figure 6.3 in chapter § 6).
2.
Cut the cable of the faulty sensor to facilitate removal.
Remove the two screws fixing the sensor block to the
cover (figure on the right).
3.
Insert the new sensor in the hole shown by the arrow in
the figure on the right.

Table of Contents