INFICON Software Operation | 9
074-594-P1H Micro GC Fusion Operating Manual 165 / 319
Method Parameter Unit Description
Temperature profile N/A
Click on the icon to display the column
temperature profile.
Sample pump time s Duration that the sample pump pulls sample into the
instrument
Continuous N/A Disables sample pump and allows sample to
continuously flow through the injector, assuming
positive sample pressure
NOTE: Not available with backflush modules
Sample inlet heater
disabled
N/A Disables the sample inlet heater
Sample inlet heater
temperature
°C Temperature set point of the sample inlet
Low start
temperature
N/A Starts the column fan to allow Micro GC Fusion to
start the run at a lower temperature
Sample conditioner
enable heater
N/A Enables the sample conditioner (if installed)
Sample conditioner
purge time
s Duration that the sample conditioner is purged (if
installed)
Sample conditioner
flow rate
N/A Sets the sample conditioner purge flow rate to low or
high (if installed)
Column Length Maximum Ramp Rate
8 m 5°C/s
10 m 4°C/s
> 10 m 2.5°C/s
Micro GC Fusion is equipped with the Fusion Auto-Sensing Technology,
known as
FAST Enabled
, automatically adjusts the gain of the detector during
a sample run, providing fast and accurate analyses of gases within a large
concentration range. Prior to being
FAST Enabled
, the instrument provided
standard or high sensitivity settings to detect high ppm to percentage and low
ppm to high ppm components, respectively.