EasyManuals Logo

Inficon Micro GC Fusion 2-Module User Manual

Inficon Micro GC Fusion 2-Module
319 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #214 background imageLoading...
Page #214 background image
9 | Software Operation INFICON
214 / 319 074-594-P1H Micro GC Fusion Operating Man-
ual
6
Type in the
Compound Name
for the peak.
7
The left RT(s) and right RT (s) columns define the retention time window. Edit
the retention time start and end points. The retention time window displays as a
yellow shaded bar surrounding the peak.
8
Type the concentration of the compound as listed on the calibration gas standard
certificate of analysis. Click
mole%
under
Concentration
to change units to
ppm
,
and vice versa. An area count (in μVs) is generated in the
Area
column. The
Fit
Type Linear Forced Zero
populates by default.
9
Click on the icon to display the calibration curve. A goodness of fit (R
2
) value
displays. The slope of the calibration curve corresponds to the compound’s
response factor.
10
Repeat steps 4 to 9 for all compounds.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon Micro GC Fusion 2-Module and is the answer not in the manual?

Inficon Micro GC Fusion 2-Module Specifications

General IconGeneral
BrandInficon
ModelMicro GC Fusion 2-Module
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals