9 | Software Operation INFICON
218 / 319 074-594-P1H Micro GC Fusion Operating Man-
ual
ð Average Replicates - Select multiple runs from the calibration gas and click
to display a window. Click to confirm you want to average the areas of
the individual compounds in the calibrated method. The runs used in average
replicates appear with a icon next to each run.
4
The
Area (μV)
column will automatically populate with the new area counts for all
compounds in the run. The name of the calibration will be replaced with the
currently selected run. When using average replicates, one of the data files used
in the calculation will be used as the reference chromatogram for the calibration.
5
If using a new calibration gas standard, type the
Concentration
of the compound
as listed on the calibration gas standard certificate of analysis.
6
Click the icon to save the new calibration to the method. A window will
prompt the user to save the calibration. To override the current method, use the
original method name. An icon will indicate that there are unsaved changes.
7
Click the icon.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon Micro GC Fusion 2-Module and is the answer not in the manual?

Inficon Micro GC Fusion 2-Module Specifications

General IconGeneral
BrandInficon
ModelMicro GC Fusion 2-Module
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals