9 | Software Operation INFICON
222 / 319 074-594-P1H Micro GC Fusion Operating Man-
ual
8
Click the icon to save the new calibration to the method. A window will
prompt the user to save the calibration. To override the current method, use the
original method name. An icon will indicate that there are unsaved changes.
9
Click the icon.
10
A message will prompt the user to reload the method and apply the calibration.
11
To reprocess the calibrated run with the new information, click the icon. A
confirmation message will display
Are you sure you want to reprocess the X
selected run(s)?
Click the icon to reprocess the run. To apply the new
calibration to prior runs, select the desired runs from the list before clicking the
icon.
12
A confirmation message will display.
13
Click the icon to exit the calibration window.
14
Click the updated method preset to reload the method.